| Applications |
Steam Condensate, Boiler Feed, Normally Clean Water
|
| Hydrocarbons |
Fuel Oil, Crude Oil, Other Fuels and Oils, Phenol, Styrene,Most Heat Transfer Fluids, Aromatic Chemicals, BTEX |
| Detection Range: Fuel Oil, Crude Oil |
5 ppb in Steam Condensate - 500ppm |
| Low Detection Limit: Other Hydrocarbons |
Depends on the Type, Generally < 50ppb |
| Dimensions |
20" x 16" x 11" (50.8cm x 40.6cm x 27.9cm) |
| Weight |
65 pounds (29.5kg) |
| Local Display |
Yes, PPM, PPB, Or Raw Signal |
| Controls |
External 316 SS Touch Pad |
| Power Requirements |
90 - 240 VAC, 2A, (24 vdc Optional) |
| Plumbing Requirements |
Feed: 3/8" or 1/2" Tube, Return: 3/8"or 1/2" Tube |
| Sample Pressure |
<150psig, 10barg (For higher pressures consult factory) |
| Sample Temperature |
100°C, 212°F (For higher temperatures consult factory) |
| Required Air Supply General Purpose |
None |
| Required Air Supply Zone I, Zone II |
15 scfh |
| Ambient Operating Temperature |
32°F - 121°F, 0°C - 49°C (Heater or Vortex Cooler Optional) |
| Flow Cell Characteristic |
Closed Tube with Internal Cell Condition Sensor |
| Pressurized Sample Return |
Yes |
| Operational Principle |
UV Fluorescence |
| Calibration Stability |
+/- 10% over 6 Months |
| Response Time |
3 seconds full scale spike change, continuous reading |
| Calibration |
Blank / Standard Addition or Correlation to Standard Methods |
| Reagents |
None |
| Alarms |
4: Early Warning / High Alarm / Cell Condition / CIP |
| Alarm Contacts |
Dry Contact AC/DC
|
| Outputs |
4-20 mA,24 vdc powered, 500 ohm impedence |
| Communication Protocols |
Optional: Hart, Mod Bus |
| Diagnostics |
Self Diagnostics, internal failure linked to alarm, Cell Condition |
| Security |
Password Protects Software, Lockable Enclosure |
| Electronics Cabinet |
316 SS |
| IP Rating |
IP W 66, NEMA 4X |
| Maintenance |
Calibration Check, Clean Flow Cell |
| Cleaning Chemicals |
Detergents and/or Dilute HCL |
| Explosion Proof Options |
Gen Purpose, (Optional C1D2, Atex Zone II, C1D1, Atex Zone 1) |
|
Sample Pump
|
Optional |
| Semi Automatic Cleaning System |
Optional |
| Data Logger |
Optional |